Decontamination Equipment Operational Parameters

Automated Washer/Disinfector compliant to HTM 2030

Wash Temp 60°C
Wash Detergent Serchem Maxim PH Plus with a PH in solution of 10.4
Disinfection Temperature 90°C for 1 minute

Pre-vacuum Steam Sterilizer compliant to HTM 2010

Sterilization plateau 134°C to 137°C for between 3 and 3½ minutes